IMJ00503 V3E.DOC Page 19 of 40 19.02.2009
4.5.9.4. Configuration Data
For input and output data, the following configurations are available:
mandantory
class 1 D1 hex 2 words inputs consistency
class 2 F1 hex 2 words of input data, 2 words of output data for preset
value, consistency
optional
class 1 D0 hex Not implemented in LMCJ0043
class 2 F0 hex Not implemented in LMCJ0043
4.5.9.5. Cyclical Data Exchange – Input (Slave > Master)
The LMC J 0050 3 outputs position data which are mathematically signed. The SF (scale factor)
parameter can be used to invert a given mathematical sign. Resolution is also defined via SF.
Octets in a telegram are arranged in a Profibus compliant manner (big endian), i.e. the MSB
always comes first and the LSB is always the last one.
Octet Bit Type Output
1….4 signed 32 Position datas from Encoder
4.5.9.6. Cyclical Data Exchange – Output (Master > Slave)
The most significant bit in a preset value (bit 32) defines the validity of that preset.
Octet Bit Type Output
1….4 signed 32 Preset Value
Normal Mode: MSB = 0 (bit 31)
Preset Mode : MSB = 1 (bit 31)
The “Preset“ value is available for setting a current output value to a desired value. This is
achieved with the help of an internal offset M
offset
. By setting of bit 31, this offset value can be
changed. The following assignments are valid:
M
DataEx
value transported to the Profibus in cyclical data exchange mode
M
Laser
reading determined by the laser
M
offset
intern calculated offset value
M
Preset
value transferred with “Preset“ function
Cyclical calculation of: M
DataEx
= M
Laser
+ M
offset
The value for M
offset
is not permanently stored in the LMC J 0043, i.e. it will be lost on
turning power off. The offset value can also be stored as a parameter in Octet 32 ... 35.
If bit 31 of M
Preset
is set, M
Offset
will be calculated in such a manner that the following
equation is met: M
Preset
= M
Laser
+ M
offset
.
The new offset value can be read in the diagnostic data as Octet 30 ... 33.